MODEL
4310 CONDUCTIVITY METER
OPERATING MANUAL

549 007/REV A/3-96
SAFETY
Please
read this information carefully prior to installing or using this equipment.
1.
The unit described in this manual is designed to be operated only by trained
personnel. Any adjustments, maintenance and repair must be carried out as
defined in this manual, by a person qualified to be aware of the hazards
involved.
2.
It is essential that both operating and service personnel employ a safe system
of work, in addition to the detailed instructions specified in this manual.
3.
References should always be made to the Health and Safety data supplied with
any chemicals used. Generally accepted laboratory procedures for safe handling
of chemicals should be employed.
4.
If it is suspected that safety protection has been impaired in any way, the
unit must be made inoperative and secured against any intended operation. The
fault condition should immediately be reported to the appropriate servicing
authority.
MODEL 4310
CONDUCTIVITY
METER
OPERATING MANUAL
CONTENTS
SECTION 1 INTRODUCTION
Instrument
Description 1.1
Instrument
Specification 1.2
SECTION 2 INSTALLATION
Unpacking 2.1
Installation 2.2
Displays 2.3
Controls 2.4
Inputs/Outputs 2.5
SECTION 3 OPERATION
Preparation
of Standards 3.1
Calibration:
With
Known Cell Constant 3.2
With
Standard Solution 3.3
Calculating
Cell Constant 3.4
Sample
Measurement 3.5
Good
Practice Guidelines 3.6
SECTION 4 MAINTENANCE
General 4.1
Probe
Replatinization 4.2
SECTION 5 OPTIONAL
ACCESSORIES
Optional
Accessories 5.1
SECTION 6 INTERFACING
SECTION 1
INTRODUCTION
1.1 INSTRUMENT DESCRIPTION
The
Model 4310 is a general purpose Conductivity Meter with auto ranging from 0.01mS to 99.9mS. Further expansion of these
ranges is possible using the optional X 10 and X 0.1 cells .
Calibration
can be easily performed by adjusting the digital display to the value of the
cell constant indicated on the pre-calibrated cell supplied. Readings are
automatically temperature compensated over the range of 0 to 100°C when using
standard cells.
Selection
of reference temperatures of 18, 20 or 25°C can be made via a dip switch
located on the rear panel of the instrument. Direct calibration can be
performed on 10mS, 1413mS, mS and 12.88mS.
1.2 INSTRUMENT
SPECIFICATIONS
RANGE RESOLUTION
0 to l999mS lmS (when used with X10
cell only)
0 to l99.9mS 0. lmS
0 to 19.99mS 0.01mS
0 to l999 mS 1mS
0 to l99.9mS 0.1mS
0 to l9.99mS 0
01mS
-10 to +105°C 0.1°C
+14 to +220°F 1°F
Accuracy: Conductivity: ±0.5% ±2 digits
Temperature: 0.5°C
ATC
Range: 0
to 100°C
Temperature
Coefficient: Variable
0 to 4.00%/°C
Excitation
Frequency: Auto ranging between 40 or 800Hz
Reference
Temperature: 18, 20 or 25°C (switchable)
Cell
Constant: 0.015
to 19.99 digitally settable
Platinising
Socket: 10V
pp 0.02Hz
Outputs: Analogue
lmV per digit
Bi-directional
RS232
Display: LCD
Power: Power
Supply
Size: 275(1)
x 240(w) x 150(d)mm
Weight: 1.7Kg
SECTION 2
INSTALLATION
2.1 UNPACKING
Remove
the Model 4310 from the packaging and ensure the items contained within the
package are as ordered.
Any shortages or damage
should be reported immediately to the Manufacturer or your local Distributor.
NOTE: Power Supply 021 033
is supplied with a moulded European plug. If this is not correct for your local
supply it should be cut off and a suitable lead connector fitted noting the
colours of the internal conductors as follows: Brown - Live Blue - Neutral
2.2 INSTALLATION
For
units supplied with the swing arm electrode holder the following assembly
instructions should be carried out:
1
Unpack the assembly and ensure the following items are present:
a)
Base block and b) Swing arm. Assemble as illustrated. The moulded pivot is a
tight push fit onto the pin.

2.
Fit the conductivity cell into the cut-out in the support block. The optional
temperature probe, if supplied, should be placed into the small hole in the
contra of the block. The cable(s) should be passed through the retaining clip
on the holder and connected to the respective socket on the rear panel.
2 549 007/REV A/03-96
2.3 DISPLAYS

1.
Main
display - provides direct readout of conductivity of samples and standards. The
display will also show OUT OF RANGE (1) symbols if the instrument is reading
outside the ranges 0 to 199.9mS or 0 to 1999mS if the X10 cell is used.
2.
CAL
- pressing ENT with the cursor beneath this option illuminates the CAL
annunciator momentarily and then calibrates the instrument to the nearest
standard solution (10mS, 1413mS, 12.88mS or 0mS) . “Err” is displayed if a cell constant
outside the allowed range is calculated and the calibration aborted.
3.
Secondary
display - provides direct readout of solution temperature in °C or °F
(selectable by DIP switch setting). Also provides readout of cell constant (K)
and temperature coefficient (%).
4.
Endpoint
detection symbol - this is displayed once a stable reading is detected, and is
maintained until the input changes.
5.
Selected
mode indicator.
6.
Cursor
- used to select/indicate required mode.
7.
Selected
unit of temperature(°C/°F), cell constant(K) and temperature coefficients(%).
8.
Measurement
unit which is being used, mS or mS.
3
549 0071REV A/03-96
2.4 CONTROLS

Switches the instrument on and
to Standby.
Print Key. Provides a printout
of the current reading with an incremental sample
number. When pressed for
the first time after a calibration, the printout will give
calibration information.
The incremental sample number will be reset after a calibrationor after changing the cell
constant (K) and/or temperature coefficient.
These keys are used to
change a parameter
. These keys are used to adjust the
cell constant or temperature coefficient when
displayed.
These keys are used to move the cursor between
the menu options.
This key is used to select the
displayed menu option.
4 549 007/REV A/03-96
2.5 INPUTS/OUTPUTS

1. 7 PIN DIN SOCKET Connection socket for the standard cell supplied with the
unit
2. 7 PIN DIN SOCKET Platinising socket
3. OUTPUT SOCKET 25 way D type socket for RS232
4. ANALOG OUT 2
x 4mm sockets. Analogue output lmV per digit
5. DIP SWITCHES Used
to select temperature units (°C or °F) and reference
temperature
18, 20 or 25°C
SW1
- switch on =°C or °F
SW
3 2 Ref. Temp.
off off IR°C
off on 20°C
on off 25°C
on on 25°C
6. POWER IN 2.1
x 5 .5mm socket allowing the power supply to be connected to
the
unit
5
549 007/REV A/03-96
SECTION 3
OPERATION
3.1 PREPARATION OF
CONDUCTIVITY STANDARDS
Suitable
conductivity standards are available commercially or these can be made up as
required from A.R. grade reagents with reference to relevant physical tables.
Method for general purpose
Conductivity Standard
Accurately
weigh out 0.746 grammes of dried A.R. grade Potassium Chloride (KCI) and
dissolve in 1 litre of good quality water. This produces a 0.01N solution with
a conductivity of 1413mS at 25°C.
Storage
This
solution must be stored in a plastic container and the air space above the
solution should be kept to an absolute minimum. The shelf life of 1 week can be
increased by storing below 4°C, but where any doubt exists about the viability
of stored solution a fresh batch should be prepared.
3.2 CALIBRATION WITH KNOWN
CELL CONSTANT
1. Connect a standard pre-calibrated cell to
the unit.
2. Select the required reference temperature by
setting the DIP switches on the rear panel to the appropriate temperature as detailed below:
SW
3 2 Ref. Temp.
off off 18°C
off on 20°C
on off 25°C
on on 25°C
3. Move the cursor beneath the K menu option
and press the
key. The secondary display will show the
current cell constant, K.
4. Use the
keys to adjust the display to indicate the
value of the cell constant as marked on the probe being used. The main display
will show the adjusted conductivity reading.
The standard X 1.0 cell has a cell constant range from 0.8 to 1.20 and
is generally calibrated to 2 decimal places. The display should be set to
indicate this figure exactly. The
standard X10 cell has a cell constant range from 8.0 to 12.0 and is generally
calibrated to 2 decimal places. The display should be set to indicate this
figure.
6
549 007/REV A/03-96
5. Move the cursor beneath the COEFF menu
option and pres’s the
key. Note that this menu option toggles the
secondary display between temperature coefficient (0 to 4%/°C) and temperature.
With the temperature coefficient displayed, the
keys can be used to adjust the temperature
coefficient.
NOTE: The unit has battery back-up to ensure all
parameters are retained in non-volatile memory. When re-using the instrument it is recommended that the
parameters previously set should be
checked to ensure the values set are relevant to the tests being carried
out.
6. Prior to use on the 20mS range the cell should be thoroughly rinsed
in deionised water, shaken to remove internal droplets, and the outside wiped
dry. Pressing the
key with the cursor beneath the CAL
menu option when the display reads less
than 2mS will illuminate the CAL
annunciator for 1 second, and then zero the display.
3.3 CALIBRATION WITH
STANDARD SOLUTION
Calibration
of the unit and cell with standard solutions will only be necessary if:
1) The cell constant is unknown.
2) The cell constant has changed due to replatinising,
wear or damage to the plates.
3) An ATC slope other than 2%/°C is required
(refer Special Calibration).
General Calibration
1)
Immerse the conductivity cell into the prepared standard.
2) Move the cursor beneath the CAL menu option.
3) Press
. She unit will calibrate to the nearest
standard.
Special Calibration
When
using conductivity as a method for determining or monitoring the concentration
of a specific species in solution, and when the temperature coefficient (slope)
is unknown, it will be necessary to prepare a standard solution of known
concentration containing only the species to be measured. The standard must be
maintained at the chosen reference temperature throughout calibration. Move the
cursor beneath the K menu option and adjust the display to read the
conductivity value of the chosen standard using the arrow keys. Samples should
also be measured at the reference temperature and, if required, the
concentration can be determined from
calibration curve or suitable physical tables (ensure these are based on
the same reference temperature used for the measurement).
7 549
007/REV A/03-96
3.4 CALCULATING CELL
CONSTANT
After
completing the calibration with standard solutions the probe should remain in
the standard solution with the main display set to the value of this solution.
Moving
the cursor beneath the K option and pressing the
key will provide a direct readout of the
cell constant. Note that pressing
with the cursor beneath this menu option
toggles the secondary display between cell constant (0.015 to 19.99) and
temperature With the cell constant displayed, the
arrows can be used to adjust the cell
constant which changes the conductivity reading using:
Displayed
conductivity = K x Measured Conductivity
3.5 SAMPLE MEASUREMENT
General
Conductivity
is a temperature dependent measurement. All substances have a conductivity
coefficient which varies from 1% per °C to 3% per °C for most commonly occurring
substances. The automatic temperature compensation on the 4310 defaults to 2%
per °C, this being adequate for most routine determinations. Conductivity
readings varying with temperature may be due to the substances under test
having a coefficient other than the typical value of 2% per °C. To eliminate
this variation it is necessary to maintain all samples at the reference
temperature by use of a thermostatic water bath or equivalent.
Adjustment
may be made by moving the cursor beneath COEFF option and pressing the
key. The reading can then be adjusted to the
required value (0.00 to 4.00) by using the
keys.
Sample Measurement
After
calibration the measurement of samples is carried out by immersing the cell in
the samples, allowing the readout to stabilise, and recording the result. The
cell should be rinsed in deionised water between each sample to avoid
contamination, shaken to remove internal droplets, and the outside wiped prior
to immersion in the next sarnple.
On
completion of sample measurement the cell should be thoroughly rinsed in
deionised water.
Storage
Short Term the cell should be immersed in deionised water to keep the
plates wetted.
Long Term the cell should be thoroughly rinsed in deionised water, the
exterior body wiped and then stored dry.
NOTE: When preparing the cell for storage the
plate area MUST NOT be wiped dry. When using a dry cell initial stability on
re-use may be impaired until the cell plates become re-wetted.
To
obtain optimum performance refer to Section 3.6 - Good Practice Guidelines.
8 549
007/REV A/03-96
3.6 GOOD PRACTICE GUIDELINES
1.
For
greatest accuracy ensure no particulate matter is suspended in the solution
under test. If necessary, filter or allow the particles to settle prior to use.
Do not allow the cell to come into contact with any sediment which may be
present.
2.
Ensure
no air bubbles are trapped in the cell area between the plates.
3.
Ensure
the cell plates are completely immersed in the solution under test.
4.
Thoroughly
rinse the cell plafes in deionised water after use, and for short term storage
immerse the cell in deionised water. Although it is not essential to store the
conductivity cell with the plates in a wetted condition, if they are allowed to
dry out completely, initial stability on re-use may be impaired until the cell
plates become re-wetted.
5.
Do
not attempt to clean the cell plates as this will remove the black
platinization and alter the calibration and accuracy of the probe. If the cell
plates become worn or damaged the plates should be re-platinised and the cell
constant recalculated.
6.
Ensure no salt deposits or particulate matter
are allowed to build up around the cell plates or on the probe body as this may
produce a conductivity path lower than that through the solution. It is
recommended that such deposits be removed by soaking the cell in deionised
water. No attempt should be made to wipe off these deposits as this may cause
damage to the cell plates.
7.
Ensure the correct reference temperature is
selected for the operating procedures being used.
9 549 0071REV A/03-96
SECTION
4
MAINTENANCE
4.1 GENERAL
The
Model 4310 is designed to give optimum performance with minimum maintenance. It
is only necessary to keep the external surfaces clean and free from dust. To
give added protection when the unit is not in use the unit should be
disconnected from the mains supply and covered with the optional dust cover
(544 008). For longer term storage or re-shipment it is recommended that the unit
be returned to the original packing case. All conductivity cells should be
thoroughly rinsed after use and stored in deionised water. Do not attempt to
clean the surface of the plates as this could result in damage to the black
platinised surface.
4.2 PROBE RE-PLATINISATION
Probe
types 027 012 and 027 028 have a black platinum film on the cell plates.
Deterioration of this film will be apparent either visually or functionally by
causing a low reading to be obtained, particularly on conductive solutions.
Should
this occur the following procedure should be carried out:
WARNING - THE CHEMICALS USED
IN THIS PROCEDURE ARE CORROSIVE. ALL NORMAL SAFETY PRECAUTIONS SHOULD BE
OBSERVED.
Solution
1.5
grammes Chloroplatinic Acid
0.015
grammes Lead Acetate
Deionised
Water
The
chemicals should be placed in a suitable container and made up to 50ml using
the deionised water.
PROCEDURE
1) Connect the probe to the rear panel
platinising socket.
2) Rinse the cell thoroughly in deionised water
to ensure it is clean and free from contamination. In extreme cases rinsing in
Chromic Acid may be necessary.
3) Completely immerse the plates in the
platinising solution.
4) The probe should be left in the solution for
a period of no greater than 10 minutes. At the end of this time the plates
should have an even black appearance.
5) Carefully remove the cell from the
platinising solution and wash in deionised water.
6)
Completely immerse the plates in a 1M Sulphuric Acid solution and leave for a
period of 10 minutes.
7)
Rinse the probe in deionised water and recalibrate the probe.
10 549 007/REV A/03-96
SECTION 5
OPTIONAL
ACCESSORIES
5.1
OPTIONAL ACCESSORIES
The
following list of items are available as optional accessories for use with the
Model 4310:
903
200 Electrode stand
with swing arm electrode holder
027
024 Temperature
probe/ATC
021
030 Power Supply (UK
Version)
021
031 Power Supply
(European Version)
021
032 Power Supply (US
Version)
021
033 Power Supply
(230V leaded)
NOTE: Power Supply 021 033
is supplied with a moulded European plug. If this is not correct for your local
supply it should be cut off and a suitable local connector fitted noting the
colours of the internal conductors as follows: Brown - Live Blue - Neutral
544
008 Dust Cover
549
009 French
Instruction Manual
549
010 German
Instruction Manual
549
011 Italian
Instruction Manual
549
012 Spanish
Instruction Manual
549
013 Portuguese
Instruction Manual
543
001 40 Column
Printer
050
002 Jensoft
542
009 Interface Cable
Kit
Conductivity Cells
027
013 Conductivity
Cell K=l
027
113 Conductivity
Cell K= 0.l
027
114 Conductivity
Cell K=1 (plastic covered)
NOTE: The cells listed below
are glass free and epoxy bodied.
027
211 Conductivity Cell with ATC K=O.1
027
212 Conductivity
Cell with ATC K=1
027
213 Conductivity
Cell with ATC K=10
Calibration Standards
025
138 1413mS Standard (500ml)
025
156 12.88mS Standard
(500ml)
025
164 84mS Standard (500ml)
025
139 10mS Standard (500ml)
025
165 1382ppm TDS
Standard (500ml)
1
1 549 007/REV A/03-96
SECTION 6
INTERFACING
Analogue
All
units are provided with 2 x 4mm sockets, marked as ANALOG OUT, on the rear
panel. An analogue output voltage of lmV per least significant digit is
available from these sockets.
;
RS232
The
Bi-directional RS232 interface is available on the rear panel 25 way D type
connector.
The
connections are as follows:
TXD
2 - INPUT TO 4310
RXD
3 - OUTPUT FROM 4310
RTS
4 - LINKED TO CTS
CTS
5 - LINKED TO RTS
DSR
6 - OUTPUT FROM 4310
DCD
8 - OUTPUT FROM 4310
DTR
20 - INPUT TO 4310 (must be
active)
GND
7
Suggested
interconnections are detailed below:
4310
IBM PC XT (25 way “D”)
TXD 2 2 TXD (From PC)
![]()
RXD 3 3 RXD (ToPC)
RTS 4 4 RTS (From PC)
CTS 5 5 CTS
(ToPC)
DSR 6 6 DSR (ToPC)
DCD 8 8 DCD (To PC)
DTR 20 20 DTR
(FromPC)
GND 7 7 GND
4310 IBM PC XT (9 way “D”)
TXD 2 3 TXD (From PC)
RXD 3 2 RXD (To PC)
RTS 4 7 RTS
(From PC)
CTS 5 8 CTS
(To PC)
DSR 6 6 DSR
(To PC)
DCD 8 1 DCD
(To PC)
DTR 20 4 DTR (From PC)
GND 7 5 S GND
12 549 007/REV A/03-96
INTERFACING (continued)
The
RS232 communications parameters on the computer or printer need to be set to
match those of the Model 4310, as detailed below:
1200 Baud
7 Data Bits
Odd Parity
1 Stop Bit
The
Model 4310 supports both hardware (DTR/DSR) flow control and software XON/XOFF
flow control.
Pressing
the PRINT key outputs from the RS232
interface.
Sending
an ASCII “D” to the 4310 causes a printout ofthe current displayed reading plus
sample number.
Sending
an ASCII “C” causes a printout of the last calibration parameters. (same as
printout after performing a calibration).
13 549 007/REV A/03-96