MANUAL No.
154

INSTRUCTIONS FOR THE
1451
SOLUTION CALORIMETER
The
unit of heat used in this Manual is the thermochemical calorie, which is equal
to 4.1840 absolute joules.
ASSEMBLY INSTRUCTIONS
Unpack
the calorimeter carefully and check the individual parts against the packing
list. If shipping damage is discovered, report it immediately to the delivering
carrier. Handle the glass sample cell and the thermistor probe with care as
these parts are fragile and easily broken. Set the calorimeter on a bench or
table in a location that is free from drafts and protected from sources of
radiant heat. Temperature changes in the room should be minimal.
Attach
the motor to the rear of the calorimeter case using the two mountings screws
which are provided. Set the cover on the calorimeter air can; drop the geared
drive belt over the motor and stirrer pulleys; connect the motor to a 115/ 120
volt, 50/60 Hz grounded outlet and start the motor. The drive system should run freely. Although the belt may appear
to be unusually loose, it is intended to operate under very light tension to
minimize friction in the stirrer bearing. The gearing on the belt and pulleys
will prevent slippage.
Turn
the selector switch on the thermometer panel to the OFF position; then connect
the power cord to a 11S/120 Volt, 50/60 Hz grounded outlet. Plug the lead wire from the thermistor probe
into the jack on the thermometer panel and lay the probe on the table near the
calorimeter. Handle the probe very carefully as the glass tip will break easily
and, if broken, it cannot be repaired.
The
unit of heat used in this Manual is the thermochemical calorie, which is equal
to 4.1840 absolute joules
Recorder
Requirements.
Instead of reading temperatures directly from the thermometer bridge with the
aid of a null indicator, it will be best in most applications to operate this
calorimeter with a potentiometric strip chart recorder to produce a thermogram
showing the temperature changes which occurred during a run. The recorder to be
used with this calorimeter should have an internal impedance of 100 ohms or
more, and it should have plotting ranges of 10, 100 and 1000 millivolts full
scale. Most laboratory recorders meet these requirements.
Connect
the recorder
to the polarized terminals at the top of the thermometer panel and plug the
power cord from the recorder into a 115/120 volt, 50/60 Hz grounded outlet. Run
a ground wire from the grounding terminal on the panel to a similiar terminal
on the recorder. If there is no grounding terminal on the recorder it may be
necessary to connect the ground wire to the instrument case but, in doing so,
do not introduce a ground loop circuit. Good grounding is very important, not
only as a safey measure, but also to ensure satisfactory thermometer and
recorder performance. If there is any question about the reliability of the
ground connection through the power cord, run a separate wire from the
grounding terminal on the calorimeter to a good earth ground.
Before
starting to use the calorimeter for the first time it will be well to make a
dry run with the calorimeter completely assembled but with no liquid in the
Dewar and no sample in the rotating cell. This will give the user an
opportunity to become familiar with the individual parts of the calorimeter and
the manner in which they must be handled. Detailed instructions are given in
the operating sections which follow.
CALORIMETER OPERATIONS
Sample
Size. The
rotating sample cell will hold up to 20 ml of liquid sample or a solid sample
weighing up to one gram, More than one gram of solid may be used in some cases,
but smaller samples are preferred so that the heat capacity and ionic strength
of the system will not change significantly when the reactants are mixed. The
Dewar must be filled with not less than 90 ml and not more than 120 ml of
liquid to properly cover the rotating cell.
Filling
the Dewar.
It is best to lift the Dewar out of the air can during the filling operation.
The liquid to be placed in it can be measured volumetrically, or the Dewar can
be placed on a solution or trip balance and filled by weight. After filling the
Dewar set it in the air can and gently push the spacer ring down as far as it
will go.
Loading
a Solid Sample. Solid samples should be suitably ground so that they will dissolve
quickly or mix uniformly with the liquid in the Dewar. Place the 126C Teflon
dish on an analytical balance and weigh the sample directly into the dish. Be
careful not to drop any of the sample into the push rod socket. After the final
weighing, set the dish on a flat surface and carefully press the glass bell
over the dish to assemble the cell. Do not grasp or press the thin-walled glass
stem during this operation; it is fragile and will break easily. Instead, grasp
the bell and press it firmly onto the dish. Then lift the cover from the
calorimeter and attach the cell to the stirring shaft by sliding the plastic
coupling onto the shaft as far as it will go and turning the thumb screw finger
tight.
Hold
the cover in a horizontal position and lower it carefully until the bottom of
the rotating cell rests on a firm, flat surface; then insert the push rod
through the pulley hub and press the end of the rod into the socket in the 126C
sample dish.

Figure
1.
Loading
a Liquid Sample. Liquid samples can be measured into the rotating cell either by volume
or by weight. Best precision is obtained by weighing, but filling from a
volumetric pipet may be adequate in come cases. Set the 126C Teflon dish on a
flat surface and press the glass bell over the dish, handling the glass
carefully as described above. If the sample is to be weighed, tare the empty
cell on a laboratory balance; insert a pipet through the glass stem and add the
liquid; then reweigh the cell. Attach the cell to the stirring shaft and insert
the push rod.
Install
the thermistor probe in the cover opening and press the bushing firmly into place to anchor
the probe in its proper position. Handle the cover carefully after installing
the probe since the glass stem will break easily.
Lower
the cover assembly with the cell and thermistor probe into the Dewar and set the cover in
place on the air can; then drop the drive belt over the pulleys and start the
major.
Combining
the Reactants.
Each test in a solution calorimeter can be divided into three distinct time
periods: (1) a preperiod during which the reactants are allowed to come
to thermal equilibrium; (2) a reaction period during which the reactants
are combined and an enthalpy change occurs in the system; and (3) a postperiod
during which the calorimeter again comes to equilibrium. At the end of the
preperiod, start the reaction by pressing the push rod downward to drop the
sample out of the rotating cell. This should be done quickly without
interrupting the rotation of the rod without undue friction from the finger.
Push the rod down as far as it will go; after which it should continue to
rotate with the pulley. Let the stirrer continue to run during the postperiod
until a uniform slope is established, as explained later in these instructions.

THERMOMETER
AND RECORDER OPERATIONS
The
temperature measuring system in this calorimeter consists of a thermistor probe
and a special bridge designed for use within the ten degree span from 20° to
30° C. Within this range the response of the thermometer is linear, with each
100 microvolt change in output representing a temperature change of exactly .
001° C. Thus, when the out-put signal is fed to a recorder and plotted on a 10,
100 or 1000 millivolt chart, temperatures can be read directly from the chart
without applying a conversion factor. The following values will apply when
equating changes in thermometer output to temperature changes:
100
microvolts (0. 0001 V) = 0. 001° C
10
millivolts (0. 010 V) =
0. 100°
100
millivolts (0. l00 V) =
1. 000°
1000
millivolts (1. 000 V) =
10.00°
Once
these basic relationships are understood the bridge can be balanced to a zero
output at any baseline temperature from 20° to 30° C and a recording range can
be selected to produce a full-scale trace corresponding to a temperature change
of 0. 1°, 1. 0° or 10° C. The true temperature at any point on the chart can
then be determined by adding the chart reading to the baseline setting shown by
the unit temperature switch and digital potentiometer located within the
marked boxc on the thermometer panel.
There
are five switch positions on the selector switch in the center of
the
thermometer panel, the middle three of which have adjusting knobs.
OFF ZERO
NULL CAL READ
In
the OFF position no power is supplied to the thermometer and the system is
dead. The switch should remain in this position whenever the calorimeter is not
in use.
The
ZERO control adjusts the output of the bridge to zero voltage.
The
NULL control adjusts the bridge to indicate a temperature of
exactly
20° C at zero voltage output.
The
CAL adjustment sets the full scale output of the bridge at exactly 1000
millivolts ( 1.000 V), corresponding to a temperature exactly 10° C above the
bridge null.
The
READ position is used to feed the thermometer signal to the recorder.
Calibrating
the recorder.
Most strip chart recorders can be set to produce a full scale trace with inputs
of 10, 100 or 1000 millivolts, which cover the ranges generally used for the
1451 Calorimeter. The recorder should have a chart speed selector and an
adjustment for setting the pen to a zero baseline. Any specific instructions
furnished by the recorder manufacturer should be observed when using this
equipment.
After
connecting the recorder to the thermometer bridge, balance the bridge and
calibrate the recorder using the procedure described below. These steps should be taken in sequence when
using the recorder for the first time. It is not necessary to repeat these
adjustments in each subsequent run, but the settings should be checked from
time to time to be sure that they have not changed.
1.
Turn the recorder on and turn the thermometer selector switch to ZERO for a
warm up period before making any adjustments. Although the thermometer will
usually warm up in ten minutes, a longer period up to thirty minutes may be
required to reach maximum stability.
2.
Start the chart drive at a speed of one inch per minute.
3.
Move the recording pen to the zero baseline on the chart as instructed by the
recorder manufacturer.
4. Set the range switch on the recorder to 1000
millivolts (1. 000 volt) full scale. In this position the full scale of the
chart represents a span of 10° C.
5.
With the thermometer selector switch in the ZERO position, use the ZERO
adjustment on the thermometer bridge to bring the pen back to its zero
baseline.
6.
Turn the selector switch to NULL and bring the pen to its zero baseline with
the NULL adjustment.
7.
Turn the selector switch to CAL and use the CAL adjustment to move the pen to
its full scale position.
8.
Now set the unit temperature switch and the digital potentiometer to read
exactly 20. 000°C and turn the selector switch to the READ position. The pen will then move to a position on the
chart indicating the temperature sensed by the thermistor.
Example:
If
the chart paper has 10 major units in its ruling and the recorder is set at
1000 millivolts (10. 00°) full scale, each major unit on the chart represents
1°C. Therefore a reading of 4. 52 units on the chart indicates a temperature of
24. 52° C in the calorimeter (20° baseline + 4. 52° on the chart scale).
Better
precision
can now be obtained by changing the baseline setting and increasing the
sensitivity of the recorder after it is known that the temperature being
measured is near 24. 52°. Move the temperature setting on the bridge to exactly
24. 000°, then change the range selector on the recorder to 100 millivolts (1.
000°) full scale. If the pen then moves to 5. 23 major divisions on the chart,
the temperature in the calorimeter is 24. 523° (24° baseline + . 523° on the
chart).
Or,
for best precision, set the temperature dials to exactly 24. 520°
and
change the range selector on the recorder to 10 millivolts (0.100°)
full
scale. Now use the recorder as a null indicator and adjust the digital
potentiometer
to bring the pen back to the zero baseline. If the digital
meter
then reads .525, the temperature in the calorimeter is 24. 525° C.
PRODUCING
THE THERMOGRAM
Choosing
the Range.
Before starting a run, try to estimate the total
energy
change involved in the experiment so that the recorder can be set
in
a range which will produce the best thermogram. The following set-
tings
are recommended:
Up
to 10 calories,
set at 0.01 V (0.100°) full scale.
10
to 100 calories, set at 0.10 V (1. 000°) full scale.
100
to 1000 calories, set at l. 00 V (10. 00°) full scale.
If
the sign and magnitude of the enthalpy change calmot be estimated before
starting an experiment, set the recorder at 0.1 V (1. 000°) full scale and make
a trial run starting with a baseline in the middle of the chart. The
temperature change observed in this experiment can then be used as a guide for
setting the recorder to a different range and baseline to produce a better
thermogram in a subsequent run.
Resetting
the Pen. If
the reaction to be examined is expected to be endothermic, the pen must be
raised to a higher position in order to record a temperature drop during the
reaction. To raise the pen, simply turn the unit temperature switch or the
digital potentiometer to a lower setting. This change can be made at any time
without affecting the range and calibration of the recorder.
In
exothermic experiments, a baseline adjustment may be desirable at this time if
the temperature in the calorimeter has changed significantly during the initial
equilibration period.
Beginning
the Trace.
The liquid in the Dewar and the sample in the rotating cell must reach thermal
equilibrium; and the recorder must trace a straight line for at least 3 or 4
minutes before starting the reaction. To minimize this equilibration period,
the reactants should be at approximately the same temperature when they are
placed in the calorimeter. This is particularly important when working with
reactions which produce low enthalpy changes. In such cases any temperature
difference between the two solutions in a liquid-liquid system should not
exceed 0. 2° when the calorimeter is loaded. The calorimeter should then be
allowed to run for about 15 minutes before starting the trace. Solid-liquid
systems will usually come to equilibrium within a shorter period.
Completing
the Trace.
Having established the initial drift, start the reaction by depressing the push
rod to open the rotating cell. This will produce a distinct temperature change
which will soon taper off to a uniform drift as the calorimeter again comes to
equilibrium. Continue the trace until the drift line becomes straight and
remains straight for at least three minutes. Usually this condition will be
reached within ten minutes or less after starting the reaction.
At
the conclusion of the test, stop the recorder; lift the pen and turn the
thermometer selector switch to ZERO. Remove the chart from the recorder and
mark it to identify the run and the reactants involved. Also, write in the
baseline temperature and show the recorder range setting for this run.
Emptying
the Calorimeter. Stop the calorimeter motor; raise the cover carefully and wipe any
excess liquid from the parts that were immersed in the Dewar. Remove the
thermistor probe from the cover and remove the sample dish from the end of the
push rod; then remove the rod and release the glass cell from the drive shaft.
Lift the Dewar out of the air can and empty it; then wash and dry all wetted
parts carefully.
READING THE
THERMOGRAM
In
order to determine the net temperature change produced by the reaction it is
necessary to locate a point on the thermogram at which the temperture reached
63 per cent of its total rise (or fall). This can be done easily using the
graphic procedure which is described below, although other variations of this
method can be used as well.
1. Place a straight edge
over the preperiod drift line and extend this line well past the point at which
the cell was opened to start the reaction.
2. Moore the straight edge
to the postperiod drift line and extrapolate this line backward to the time
when the cell was opened. If there are fluctuations in the drift lines due to
noise or other variations in the signal, use the best average when drawing
these extrapolations.
3. Using a centimeter scale,
measure the vertical distance, R, between the two extrapolated lines at a point
near the middle of the reaction period.
4. Multiply the distance, R,
by 0. 63, then
5. Set the zero end of the
centimeter scale on the extrapolated preperiod drift line and move the scale
along this line to locate a vertical intercept with the thermogram which is
exactly 0. 63R above the preperiod drift line. Draw a vertical line through
this point to intercept both drift lines.
6. Read the initial
temperature, Ti s and the final temperature, Tf,
at the points of intersection with the drift lines and subtract to determine
the corrected temperature rise, DT
DC = Tf - Ti
Although the thermogram
shown on page 13 to illustrate this graphic method is taken from an exothermic
reaction, the same steps can be used to establish the corrected temperature
change for an endothermic reaction.
If
it was necessary to reset the pen during the test, the graphic procedure
can
still be used by taking into account the two different baseline tempera-
tures
when reading the intercepts. However, in such cases it usually
will
be desirable to re-run the experiment using a different baseline or
a
higher span on the recorder to produce an unbroken thermogram.

CALCULATING THE ENERGY CHANGE
The
energy change, Q. measured in this calorimeter is calculated by multiplying the
net corrected temperature change, DTc by the energy
equivalent, e, of the calorimeter and its contents.
Q
= ( D Tc) (e)
If DTc is measured in degrees C and e
is expressed in calories per degree C, Q will be reported in calories.
The
energy equivalent, e, is determined by a calibration procedure which is described below
under Standardization.
The
change in enthalpy, DH, at the mean reaction
temperature is equal to -Q divided by the amount of sample used in the
experiment, expressed either in moles or grams.
DHT = -Q/m
where
T is the temperature at the 0. 63R point on the thermogram.
Enthalpy
values are
usually expressed in kilocalories per mole. Procedures for converting enthalpy
changes, DH. to thermodynamic standard
conditions and for using DH in other computations can
be obtained from thermodynamics or thermochemistry textbooks, or from
literature references.
STANDARDIZING
THE CALORIMETER
As
explained above, in order to calculate the energy change, Q. involved in a
reaction, the energy equivalent, e, of the calorimeter and its contents must be
known. Values of e are determined by running several calibration experiments in
which the calorimeter is operated in the usual manner but with reactants which
release (or absorb) a known amount of energy. The energy equivalent is then
calculated by dividing the known energy input, QE, by the corrected
temperature rise, DTc .
e =
QE/DTc
Standardization
with TRIS. A sample of tris(hydroxymethyl)aminomethane, commonly called TRIS,
is furnished with the 1451 calorimeter to provide a reliable standardizing
reagent. TRIS is furnished as a dry powder which can be used directly from the
bottle as supplied without further preparation, but undue exposure to air and
moisture should be avoided in order to preserve the integrity of the standard.
For
standardizing the 1451 calorimeter, solid TRIS can be dissolved in dilute
hydrochloric acid in a controlled reaction for which the amount of heat evolved
is well established. In the recommended standardization procedure described
below, O. 5 gram of TRIS is dissolved in 100 ml of 0.1 N HC1 to evolve 58. 738
calories per gram of TRIS at 25° C. This is the procedure:
1.
Tare the Dewar on a solution or trip balance and add exactly 100. 00 ± 0.05 grams of 0.100 N HC1.
2.
Weigh 0. 50 + .01 gram of TRIS into the 126C Teflon dish on an analytical
balance to an accuracy of ± . 0001 g.
3.
Assemble the rotating cell; place it in the calorimeter and start the motor.
4.
Let the calorimeter come to equilibrium; then set the recorder
at
0. l0V(1. 000° C) full scale; set the baseline at the bottom of the chart
for
an exothermic reaction and trace a thermogram as previously described.
5.
Analyze the thermogram to determine the net corrected temperature rise, DTc.
6.
Calculate the known energy input by substituting in the equation:
QE = m[58. 738 + 0. 3433(25 - T .63R)]
where, QE = the energy input in calories
m = the weight of TRIS in grams
T .63R = the temperature at the 0. 63R point on the
thermogram
The
term: 0. 3433(25 - T.63R)
adjusts the heat of reaction to any temperature above or below the 25° C
reference temperature.
7.
Calculate the energy equivalent of the calorimeter and its contents by
substituting in the equation:
e =
QE/ DTc
where
e will be expressed in calories per °C.
8.
Determine the energy equivalent of the empty calorimeter by subtracting the
heat capacity of the 100 g of 0.1 N HCl from e, as follows:
e’ = e - (100. 00)(0.99894)
where, e’
= the energy equivalent of the
empty calorimeter in calories per °C.
100. 00 = the weight of 0.100 N
HCl in grams
0.99894 = the specific heat of 0.1 N HCl at 25° C
Example:
A
standardization reaction involving:
0.5017 grains of TRIS, and
100.00 grams of 0.100 N HCl
produces
a net corrected temperature rise of:
DTc = 0.244° C
with
0.63 rise, T.63R , at 24.301 °C
In
this reaction the known energy input is:
QE =
0.5017 [58.738 + 0.3433(25 - 24.301)]
=
29. 589 calories
The
energy equivalent, e, of the calorimeter and its contents is then computed:
e =
29.589/0.244
=
121.27 calories/°C
The
energy equivalent, e’, of the empty calorimeter is then computed:
e’ = 121.27- (100.00)(.99894)
=
21. 38 calories/ ° C
EXAMPLE A - An Exothermic
Reaction
Problem:
Determine the change in enthalpy for solid sodium sulfate, Na2SO4,
when dissolved in a 5 gram/liter aqueous solution of barium chloride, BaCl2·2H2O .
Na2SO4 m = 0.1458 grams
Ba++ solution = 100. 00 grams
Corr. temp. rise DTc =
.042 C
(from chart, p. 20)
T.63R = 24.885 C
Energy
equivalent e =
121.46 cal/ C
Energy
evolved Q =
(DTc)(e)
=
(.042)(121.46)
=
5.1013 calories
Enthalpy
change DHT = -Q/m
= -5.1013/0.1458
=
-34. 99 cal/g @ 24. 885° C
Or,
multiplying by 142.04 (the molecular weight of Na2SO4)
DHT
= (- 34. 99)(142.04)
= -4. 970 Kcal/mole @ 24.885 C

EXAMPLE B - An Endothermic
Reaction
Problem:
Determine the heat of solution of solid potassium nitrate,
KNO3,
when dissolved in water.
KNO3 m = 0. 7180 gram
Distilled water = 100. 00 grams
Corr. temp. rise DTC
=
-. 508° C
(from chart, p. 2 2)
T.63R =
25.400 C
Energy equivalent e
= 121. 46 cal/°C
Energy evolved Q
= (DTc )(e)
=
(-. 508)(121. 46)
=
-61 . 70 calories
Enthalpy change DHT =
-Q/ m
= -(-61. 70)/. 7180
= +85. 94 cal/g @ 25. 400° C
Or,
multiplying by the molecular weight of KNO3
DHT = (85. 94)(101. 10)
=
8. 6 9 Kcal / mole @ 25.400 ° C

GENERAL INSTRUCTIONS
• For best results, always
operate the calorimeter as close to room temperature as possible .
•
When working with two liquid reactants, both liquids should be adjusted to
nearly the same temperature before they are loaded into the calorimeter. This
can be done in various ways, such as: storing the two flasks in a constant
temperature cabinet or on a heavy alum-inum plate, or by immersing the two
flasks in the same water bath.
•
Be consistent in technique. For example, if the Dewar is filled by weight
during standardization runs, weigh the liquid sample into the Dewar during
experimental runs.
•
Do not obstruct the breather hole in the air can or the passage to it through
the surrounding insulation.
•
Turn on the recorder and the thermometer and let them warm up for at least ten
minutes (and preferably up to thirty minutes) before starting the first run in
a series.
•
Clean the Dewar and the sample cell thoroughly after each run to prevent the
accumulation of contaminants.
REFERENCES
Since
this is primarily an operating manual, the user may want to consult other
references for additional information on calorimetric and thermochemical
theory. Suggested references include:
1.
Dickinson,
H. C., Combustion Bomb Calorimetry, Bulletin of the U.S.
Bureau of Standards, 11, 189 (1915).
2.
Eatough,
Christensen and Izatt, Experiments in Thermometric Titrimetry
and Titration Calorimetry, Brigham Young Univer-sity Press,
Provo, Utah (1974).
3.
Lewis
and Randall, revised by Pitzer Brewer, Thermodynamics, 2nd Edition,
McGraw-Hill Book Co., New York (1961).
4.
Rossini,
F. D., (Ed. ), Experimental Thermochemistry, (IUPAC),
Interscience Publishers, Inc., New York (1956).
5.
White,
Walter P., The Modern Calorimeter, The Chemical Catalog
Co., New York (1928).
MAINTENANCE INSTRUCTIONS
• Examine the rotating cell
periodically for leaks by closing the cell and submerging the empty bell in a
beaker of water for a period of time comparable to a complete run in the
calorimeter. If water migrates into the sealing area between the Teflon dish
and the glass bell, the seal is deteriorating and the 126C Teflon dish should
be replaced. 0
•
If the plastic coupling becomes detached from the glass stem of the sample
cell, the two parts may be rejoined by applying a small amount of a slowcure
epoxy adhesive to the top of the glass stem.
• After one hundred hours of
operation, or once each year, apply one drop of instrument oil or household oil
to the motor bearing, and one drop of oil to the bronze stirrer shaft bearing.
Be sure to remove any excess oil that may appear at the bottom of the shaft.