Flinn QuickStartsTM
Direct-Connect Temperature Probe (TC1101)
The Connections:
1. Connect the
Direct-Connect Temperature Probe to Port 1 of the Serial Box Interface.
2. Connect the Serial
Box Interface to the COM I or COM2 port of the computer using the cable
provided. (Use the 9-pin to 25-pin adapter if necessary.)
3. Plug the Serial
Box Interface into an electrical outlet using the AC adapter.
The Set-up/Customizing Data Logger:
1.
Open the Data Logger program. Click on Start,
Program, Vernier, Logger Pro.
Note: If a message
box reading “Warning! Cannot find the interface box” appears:
a) Check all
connections,
b) Determine which
port (COM1 or COM2) is accepting the cable from the Serial Box
Interface, and
c) Click on the appropriate
button (“COM1” or “COM2”).
A default graph call
untitled- 1” will appear.
The right mouse button Is used only to access Data logger
menus. Use the left mouse button for all other operations.
2.
Under the “File” menu, choose “Open... ”. A new window will appear, listing
the various experiment files for Data Logger.
3.
From the list, double-click on “Temperature Probes” choose “Direct Connect Temperature Probe” to open the file for the direct
temperature measurements. You may need to scroll the list by clicking on the
left mouse button and dragging the cursor across the list to find these files.
5. The
“Setup” menu can be used to specify
Sensor [Sensor setup, Calibration, and Details(units can be specified
here)]
Data Collection [Mode, Sampling (Rate)]
and
Interface details (Com port).
6. You do not need to
customize the axes for the introductory activity. However, if you would like to
customize the axis:
a) Double-click
anywhere on the graph. This cases the “Axes” dialog box to appear.
b) Click and hold the
cursor on the y-axis label (which currently reads
“Temp 1 & Temp
2”). Move the cursor to “Temperature 1” to collect data using Port 1 only. If
you are using temperature probes in both ports, you do not have to do this.
c) To change the
y-axis temperature range, simply Double-click inside the upper and/or lower
limit value box and enter the desired value.
d) The x-axis should
already be in units of “Time.” You may change the values in the upper and/or
lower limit value boxes in the same manner as you did for the y-axis above.
You are now ready to
collect data. Simply click “Collect” when you are ready to begin; click “Stop” to
discontinue data collection.
Introductory Activity:
Calibration:
1. Open the “Direct
Connect Temperature probe” file and load the calibration file, as described above.
2. Prepare a water
bath by adding 300 ml of water to Dewar.
3. Place the Standard
Temperature Probe in the water bath.
4. Click on ”Collect”.
5. Click “Stop” to
discontinue data collection.
6. Heat a water bath to
boiling!
7. Place the Standard
Temperature Probe in the water bath.
8. Click on ”Collect”.
9. Use Calibration in
the “Experiment” menu to calibrate the sensor using the water baths at 0 °C and 100 °C.